Related Experiment Video
Updated: Aug 25, 2026

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Application of heated electrodes operating in a non-isothermal mode for interference elimination with amperometric
Carolin Lau1, Sabine Reiter, Wolfgang Schuhmann
1Fachbereich Chemie, Abteilung für Analytische, Technische und Umweltchemie, Universität Rostock, Albert-Einstein-Strasse 3a, 18051, Rostock, Germany.
Abstract:
Heated electrodes were applied for the non-isothermal operation of amperometric glucose biosensors based on glucose oxidase immobilised on the electrode surface by entrapment within a polymer layer. The localised deposition of the polymer film under simultaneous entrapment of the enzyme was achieved by an electrochemically induced pH-modulation in the diffusion zone in front of the electrode, thus altering the solubility of the polymer chains. This non-manual sensor preparation protocol could be successfully used for the modification of a novel indirectly heated electrode. The non-isothermal operating mode allows working at the optimum temperature of the enzyme sensors without any thermal distortion of the bulk solution. Increased surface temperature of the sensor thus accelerates transport as well as kinetic processes, resulting in an enhanced amperometric signal. In the presence of interfering compounds such as ascorbic acid, the proposed technique allows use of the diverging thermal impact on the sensing process, for different electrochemically active compounds, for a deconvolution of the amperometric signal at different electrode temperatures. A calculation method for determination of glucose in the presence of one interfering compound is presented as a basis for a calculative interference elimination.
Related Concept Videos
Amperometry: Overview
Interfacial Electrochemical Methods: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Potentiometry: Membrane Electrodes
Electrodeposition
Electrodeposition can...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...