Related Experiment Video
Updated: Aug 25, 2026

Quantification of Metal Leaching in Immobilized Metal Affinity Chromatography
Published on: January 17, 2020
Amperometric detection of heavy metal ions in ion pair chromatography at an array of water/nitrobenzene micro
S Wilke1, H Wang, M Muraczewska
1Fachbereich Chemie, Martin-Luther-Universität Halle-Wittenberg, D-06217, Merseburg, Germany.
Abstract:
A novel amperometric detector for heavy metal ions has been developed and successfully applied for ion pair chromatography. The detector is based on the electrochemical transfer of the metal ions across an array of water/nitrobenzene micro interfaces. The ion transfer is facilitated by the neutral ionophores methylenebis(diphenylphosphineoxide) and methylenebis(di- phenylphosphinesulfide). More than eight metals are separated in less than 15 min on an RP18 column using octyl sulfonate as ion pair reagent. For the heavy metals, the limits of decision are 19(Pb(2+)), 9(Zn(2+)), 9l (Co(2+)), 8(Cd(2+)) and 1.6(Mn(2+)) microg/L. The applicability of the new method for water samples is demonstrated.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Ion-Exchange Chromatography
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...