Related Experiment Video
Updated: Aug 24, 2026

Multi-analyte Biochip (MAB) Based on All-solid-state Ion-selective Electrodes (ASSISE) for Physiological Research
Published on: April 18, 2013
Sonoelectroanalysis: investigation of bismuth-film-modified glassy carbon electrodes
Craig E Banks1, Jaanus Kruusma, Michael E Hyde
1Physical and Theoretical Chemistry Laboratory, Oxford University, South Parks Road, OX1 3QZ, Oxford, UK.
Abstract:
Bismuth-modified glassy carbon electrodes have been investigated for their suitability in sonoelectroanalysis. The stability of the bismuth film to the application of ultrasound was assessed via voltammetric and atomic force microscopy (AFM) studies which revealed little ablation at powers up to an intensity of 130 W cm(-2) delivered from a 25-kHz sonic horn. Furthermore, bismuth-film-modified glassy carbon electrodes were evaluated for the sonoelectroanalytical quantification of zinc and cadmium. Detection limits of 2 x 10(-7) M and 6 x 10(-9) M respectively were found after a 60-s deposition time via an acoustically assisted deposition protocol.
Related Concept Videos
Potentiometry: Membrane Electrodes
Electrodeposition
Electrodeposition can...
Electrogravimetric Analysis: Overview
To test the completeness of the...