The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Long-term effects of different humidification systems on endotracheal tube patency: evaluation by the acoustic
Samir Jaber1, Jérôme Pigeot, Redouane Fodil
1Service de Réanimation Médicale, Hôpital Henri Mondor, Assistance Publique-Hôpitaux de Paris, Université Paris 12, France.
Background:
Accumulation of mucous secretions in an endotracheal tube (ETT) increases its resistance, and the amount of deposit may be affected by the quality of humidification and heating of the inspired gas.
Methods:
The authors assessed the impact of two humidification systems, a heated humidifier (HH) and a hygroscopic-hydrophobic heat and moisture exchanger (HME), on the ETT patency in patients selected to require mechanical ventilation for more than 48 h. This comparison was performed over two consecutive periods and used the acoustic reflection method, which characterizes the amount and site of ETT obstruction and allows estimating ETT inner volume and resistance. Measurements were performed three times a week over the period of mechanical ventilation. Comparisons were performed at mid duration and at the end of the mechanical ventilation period.
Results:
The HH was used in 34 patients, and the HME was used in 26 patients. The two groups had similar severity and duration of mechanical ventilation. At mid duration of mechanical ventilation (5.5 +/- 3.3 vs. 4.8 +/- 3.3 days; P = 0.4), no difference was observed in ETT volume and resistance between the two groups. At the end of the study period (10.5 +/- 5.8 vs. 9.6 +/- 6.3 days of mechanical ventilation; P = 0.4), ETT volume was reduced to a greater extent with HME than with HH (-3.3 +/- 2.9 vs. -5.1 +/- 2.5%; P = 0.008), and ETT resistance increased significantly more with the HME than with the HH (8.4 +/- 12.2 vs. 19.4 +/- 17.7%; P = 0.001).
Conclusion:
Prolonged use of humidification systems results in progressive reduction of ETT patency, and to a greater extent with HMEs than with HHs.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...