The NUGGET study: NIR ultra gold-gilded equivalency trial.
Nicolaus Reifart1, Marie-Claude Morice, Sigmund Silber
1Main Taunus Heart Institute, Bad Soden, Germany. n.reifart@reifart-partner.de
Summary
The gold-coated NIROYAL stent showed a smaller lumen diameter and higher restenosis rates compared to the stainless steel NIR stent. Clinical outcomes, including major adverse cardiac events (MACE), were similar between the two coronary stents.
Area of Science:
- Cardiovascular Medicine
- Interventional Cardiology
- Biomaterials Science
Background:
- Coronary artery disease necessitates the use of stents to maintain vessel patency.
- Drug-eluting stents and bare-metal stents have varying efficacy and safety profiles.
- Novel stent coatings, such as gold, are being investigated to improve outcomes.
Purpose of the Study:
- To compare the efficacy and safety of the gold-coated NIROYAL stent versus the stainless steel NIR stent.
- To evaluate the primary endpoint of minimum lumen diameter at 6 months post-procedure.
- To assess secondary endpoints including clinical events and angiographic restenosis.
Main Methods:
- A randomized controlled trial involving 603 patients.
- Patients were assigned to either the NIR stent (n=298) or the NIROYAL stent (n=305).
- Primary endpoint: minimum lumen diameter at 6 months. Secondary endpoints: clinical events and restenosis rates.
Main Results:
- The NIROYAL stent group had a significantly smaller minimum lumen diameter (1.64 mm vs 1.83 mm) at 6 months (P < 0.001).
- Angiographic restenosis rates were higher in the NIROYAL group (37.7% vs 20.6%) (P < 0.001).
- Six-month rates of major adverse cardiac events (MACE) were similar: 10.5% for NIROYAL and 7.4% for NIR.
Conclusions:
- The gold-coated NIROYAL stent demonstrated inferior angiographic outcomes compared to the stainless steel NIR stent.
- NIROYAL stent resulted in a smaller lumen diameter and increased restenosis, suggesting greater neointimal hyperplasia.
- Despite angiographic differences, both stents exhibited similar MACE rates at 6 months.