Automatic system for noninvasive blood pressure determination in rotary pump recipients
Heinrich Schima1, Herbert Boehm, Leopold Huber
1Department of Cardiothoracic Surgery and LBI of Cardiosurgical Research, University of Vienna, Vienna, Austria. Heinrich.Schima@akh-wien.ac.at
Abstract:
In patients with implanted rotary pumps, the arterial pressure pulsatility is usually far lower than in normal individuals. Depending on the remaining degree of pulsatility, cuff-based systems such as the classical Riva-Rocci-determination of arterial blood pressure and correlated sounds or pressure measurements based on cuffpressure oscillations become inaccurate or even impossible. Therefore, a system was developed which evaluates the flow in the radial artery using an ultrasound wristwatch sensor, and this additional information is used for pressure determination. A computerized data acquisition and cuff-control system based on a PC using Matlab software, a wristwatch ultrasound device, and a compressor-driven pressure cuff was set up. The cuff was controlled for automatic inflation and deflation cycles. Cuff pressure and arterial flow was recorded. Several algorithm strategies were developed, which gave data for systolic blood pressure and heart rate together with a reliability index for data quality. Finally, the new algorithms were implemented in a microcontroller system. Comparisons with invasive measurements showed excellent correlation with systolic blood pressure (mean deltaP -0.3 mm Hg, n = 28). During exercise of rotary pump patients and therefore enhanced pulsatility the difference from manual evaluation was -2.1 mm Hg (n = 18). In conclusion, adaptation of the classical cuff-pressure method with ultrasound evaluation of peripheral flow allows reliable determination of blood pressure in patients with low pulsatility resulting from implanted rotary cardiac assist pumps. By development of a wristwatch sensor and an automatic control system a robust method for daily use could be developed.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure