Reproducibility and reliability of fetal cardiac time intervals using magnetocardiography
P van Leeuwen1, S Lange, A Klein
1Department of Biomagnetism, EFMT, Bochum, Germany. petervl@microtherapy.de
Abstract:
We investigated several factors which may affect the accuracy of fetal cardiac time intervals (CTI) determined in magnetocardiographic (MCG) recordings: observer differences, the number of available recording sites and the type of sensor used in acquisition. In 253 fetal MCG recordings, acquired using different biomagnetometer devices between the 15th and 42nd weeks of gestation, P-wave, QRS complex and T-wave onsets and ends were identified in signal averaged data sets independently by different observers. Using a defined procedure for setting signal events, interobserver reliability was high. Increasing the number of registration sites led to more accurate identification of the events. The differences in wave morphology between magnetometer and gradiometer configurations led to deviations in timing whereas the differences between low and high temperature devices seemed to be primarily due to noise. Signal-to-noise ratio played an important overall role in the accurate determination of CTI and changes in signal amplitude associated with fetal maturation may largely explain the effects of gestational age on reproducibility. As fetal CTI may be of value in the identification of pathologies such as intrauterine growth retardation or fetal cardiac hypertrophy, their reliable estimation will be enhanced by strategies which take these factors into account.