Impact of patient weight and emission scan duration on PET/CT image quality and lesion detectability
Benjamin S Halpern1, Magnus Dahlbom, Andrew Quon
1Department of Molecular and Medical Pharmacology, Ahmanson Biological Imaging Center, David Geffen School of Medicine at UCLA, Los Angeles, California 90095-6942, USA.
Unlabelled:
This study was performed to prospectively evaluate fast PET/CT imaging protocols using lutetium oxyorthosilicate (LSO) detector technology and 3-dimensional (3D) image-acquisition protocols.
Methods:
Fifty-seven consecutive patients (30 male, 27 female; mean age, 58.6 +/- 15.7 y) were enrolled in the study. After intravenous injection of 7.77 MBq (0.21 mCi) of (18)F-FDG per kilogram, a standard whole-body CT study (80-110 s) and PET emission scan were acquired for 4 min/bed position in 49 patients and 3 min/bed position in 8 patients. One-minute-per-bed-position data were then extracted from the 3- or 4-min/bed position scans to reconstruct single-minute/bed position scans for each patient. Patients were subgrouped according to weight as follows: <59 kg (<130 lb; n = 15), 59-81 kg (130-179 lb; n = 33), and >or=82 kg (>or=180 lb; n = 9). Three experienced observers recorded numbers and locations of lesion by consensus and independently rated image quality as good, moderate, poor, or nondiagnostic.
Results:
The observers analyzed 220 reconstructed whole-body PET images from 57 patients. They identified 114 lesions ranging in size from 0.7 to 7.0 cm on the 3- (n = 8) and 4-min/bed position images (n = 49). Of these, only 4 were missed on the 1-min/bed position scans, and all lesions were identified on the corresponding 2-min/bed position images. One- and 2-min/bed position image quality differed significantly from the 4-min/bed position image reference (P < 0.05).
Conclusion:
LSO PET detector technology permits fast 3D imaging protocols whereby weight-based emission scan durations ranging from 1 to 3 min/bed position provide similar lesion detectability when compared with 4-min/bed position images.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...