Related Experiment Video
Updated: Aug 24, 2026

High-throughput, Automated Extraction of DNA and RNA from Clinical Samples using TruTip Technology on Common Liquid Handling Robots
Published on: June 11, 2013
Development of an integrated automation system with a magnetic bead-mediated nucleic acid purification device for
Jun-Ichi Akutsu1, Yuriko Tojo, Osamu Segawa
1Department of Biotechnology and Life Science, Tokyo University of Agriculture and Technology, Naka-cho, Koganei, 184-8588, Japan.
Abstract:
We have developed an integrated automation system for genetic analysis and gene manipulation. The system, SX-8G Plus, is equipped with an 8-nozzle dispensing unit, a thermal cycler, a cooled reagent reservoir, four tip storage racks, four microplate platforms, buffer reservoirs, an agarose gel electrophoresis unit, a power supply, a pump for exchanging electrophoresis buffer, and a CCD camera. Automation of nucleic acid extraction and purification, the most difficult step in automating genetic analysis and gene manipulation, was realized using magnetic beads with Magtration Technology, which we have previously developed for automating the handling of paramagnetic beads. Using this system, we could perform the automated separation and purification of DNA fragments by agarose gel electrophoresis starting from sample loading. The system would enable the automation of almost all procedures in genetic analysis and gene manipulation.