A surface plasmon resonance immunosensor based on the streptavidin-biotin complex
1Institute of Molecular and Biomolecular Electronics, University of Wales, Bangor, Gwynedd, UK.
Biosensors & Bioelectronics
|January 1, 1992
Abstract:
It is shown that a streptavidin monolayer immobilized onto an evaporated gold film with biotin forms the basis of a highly specific sensing element. As an example, we show that by immobilizing the biotinylated antibody sex hormone binding globulin (alpha-SHBG) to the bound streptavidin monolayer a specific sensor for the antigen SHBG is readily fabricated. The interaction between immobilized antibody and corresponding antigen is monitored by surface plasmon resonance spectroscopy and is shown to follow a classic Langmuir isotherm. Detection of SHBG at nanomolar concentrations is demonstrated.