[Erroneous X-ray dose measurements in paediatric cardiac investigations]
D G Onnasch1, F Schröder, G Fischer
1Klinik für Kinderkardiologie, Christian-Albrechts-Universität zu Kiel. onnasch@pedcard.uni-kiel.de
Abstract:
According to national and international rules the x-ray exposure of cardiac catheter examinations have to be measured. For a Philips Integris angiographic system used for paediatric heart catheterization we found severe errors in the indicated time of fluoroscopy and total dose area product ( DAP) both for fluoroscopy and radiography. Fluoroscopy times shorter than 6 seconds and DAPs smaller than 0.1 Gy cm (2) are ignored and not considered in the indicated total fluoroscopy time and the fluoroscopic and radiographic DAPs. This error leads to a systematic underestimation of the risk of radiation-induced cancer especially with infants and children.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...