Related Experiment Video
Updated: Aug 24, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Geometric and material nonlinearity in tensioned wires of an external fixator
1Micro/Nano Bioengineering Laboratory, Faculty of Engineering, Department of Biological and Agricultural Engineering, Driftmier Engineering Center, University of Georgia, Athens, GA 30602, USA. gzhang@engr.uga.edu
Objective:
To investigate the fundamental characteristics of a tensioned wire in an external fixator.
Design:
Many factors that may influence the wire performance including the change in wire geometry, material hardening and yielding, loading and unloading, and levels of pre-tension are considered in order to obtain a whole spectrum of the wire characteristics.
Background:
External fixation is widely used in the treatment of unstable fractures, limb lengthening, and congenital and pathological orthopedic deformities. With the use of tensioned wires, external fixation provides attractive features such as minimal invasiveness, maximum tailorability, and versatility. These seemingly simple wires actually fulfill a very complex duty. To be able to maximize the benefit of these wires, it is necessary to know their fundamental characteristics.
Methods:
A single wire was isolated from an external Ilizarov ring-frame, and nonlinear elastic and plastic analyses of the wire were performed using the method of finite element analysis.
Results:
The nonlinear behavior of the wire originates not only from the material hardening and yielding but also from the change in its geometry. Pre-tension reduces wire deflection, delays the onset of full-plastic deformation, and elevates the limiting plastic moment, but at the same time leads to early material hardening and yielding.
Conclusions:
An entire load-deflection curve is necessary when comes to describe the wire performance. To enhance the wire performance, further efforts should be directed to exploiting the benefit of geometric nonlinearity of the wire.
Relevance:
The findings will lead to refinement of external fixation design and provide engineering information to surgeons regarding the application of pre-tension during surgery.
Related Concept Videos
Temperature Dependent Deformation
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55 °C.
Hooke's Law
Cable Subjected to a Distributed Load
Stability of structures