Thermal assessment of 40-MHz ultrasound at soft tissue-bone interfaces
Allison S Duckett1, Angela D Reid, Lisa Leamen
1Imaging Research, Sunnybrook Health Sciences Centre, Toronto, ONT, Canada. aduckett@sten.sunnybrook.utoronto.ca
Abstract:
Tissue exposure to diagnostic ultrasound (US) can cause significant temperature rises. However, little has been reported on thermal effects of high-frequency US, and guidelines for the use of US do not necessarily apply to higher frequencies. Temperature rise induced by US biomicroscopy (UBM) was measured in phantoms containing mouse skulls and in anesthetized mice and mice post mortem, with a 50-microm K-type thermocouple. The operating frequency was 40 MHz with a free field I(SPTA) of 2.6 mW/cm(2) (B-mode) and 11.9 W/cm(2) (Doppler). Peak negative pressures were 5.22 MPa (B mode) and 7.32 MPa (Doppler), resulting in a mechanical index (MI) of 0.83 (B-mode) and 1.05 (Doppler mode). In Doppler mode, mean temperature rises of 1.80 degrees C and 1.73 degrees C were measured for proximal and distal skull phantom surfaces after a 3-min insonation. In vivo, the proximal mouse skull surface showed a mean temperature rise of 2.1 degrees C, with no statistically significant differences post mortem. Our results indicate temperature rise from insonation of bone interfaces using similar exposure parameters should not cause adverse bioeffects.