Asystole following left ventricular pacing.

Rebecca W Lai1, Kent J Volosin

  • 1University of Pennsylvania Health System, Philadelphia, Pennsylvania, USA.

Pacing and Clinical Electrophysiology : PACE
|June 11, 2004
PubMed
Summary

Pacing the left ventricle during biventricular ICD implantation can cause asystole due to the Bezold-Jarisch reflex. This reflex involves C fiber stimulation, leading to a sudden drop in heart rate and blood pressure.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Enhancing patient care by estimation and discussion of risk for ICD shock.

Pacing and clinical electrophysiology : PACE·2014
Same author

Combining shock reduction strategies to enhance ICD therapy: a role for computer modeling.

Journal of cardiovascular electrophysiology·2010
Same author

Underestimation of pacing threshold as determined by an automatic ventricular threshold testing algorithm.

Pacing and clinical electrophysiology : PACE·2006
Same author

Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results.

Circulation·2004
Same author

Experience with a dual chamber implantable defibrillator.

Pacing and clinical electrophysiology : PACE·2002

Area of Science:

  • Cardiology
  • Electrophysiology
  • Medical Devices

Background:

  • Biventricular implantable cardioverter-defibrillator (ICD) implantation is a standard procedure for heart failure.
  • Epicardial lead placement in the left ventricle (LV) is sometimes necessary for optimal device function.
  • Coronary sinus (CS) lead placement is a common approach for LV pacing.

Observation:

  • During LV epicardial pacing via the coronary sinus during ICD implantation, a sudden onset of asystole was observed.
  • Asystole is defined as a complete cessation of electrical activity in the heart.

Findings:

  • The observed asystole is proposed to be a manifestation of the Bezold-Jarisch reflex.
  • This reflex is triggered by direct stimulation of cardiac C fibers.
  • The reflex results in a significant and rapid decrease in both heart rate and blood pressure.

Implications:

  • Understanding the Bezold-Jarisch reflex is crucial for managing potential complications during ICD procedures.
  • This finding highlights the importance of careful lead placement and monitoring during epicardial LV pacing.
  • Further research may explore strategies to mitigate or prevent this reflex during cardiac interventions.