The 'fixed performance' venturi: effect of downstream pressure on outflow and FIO2
1University Department of Anaesthesia, Addenbrooke's Hospital, Hills Road, Cambridge, CB2 2QQ, UK. Gareth@garjons.demon.co.uk
Fixed performance venturi devices are sensitive to downstream pressure, affecting oxygen delivery. Even minimal pressure changes significantly reduce outflow and alter oxygen concentration, potentially below adult needs.
Area of Science:
- Medical Devices
- Respiratory Therapy
- Biomedical Engineering
Background:
- Fixed performance venturi devices are intended to deliver a specific oxygen concentration at an outflow exceeding adult peak resting inspiratory flow rates (approx. 30 L/min).
- Previous literature noted venturi device sensitivity to downstream resistance but lacked detailed investigation.
Purpose of the Study:
- To examine the impact of increased downstream pressure on the outflow and delivered oxygen concentration of standard venturi devices.
- To determine if venturi devices maintain predetermined oxygen concentrations under varying downstream resistance.
Main Methods:
- Five standard venturi devices (24-60% O(2)) were tested.
- Outflow and oxygen concentration were measured under increasing downstream pressure conditions.
Main Results:
- Venturi device outflow was highly sensitive to small increases in downstream pressure.
- A minimal increase in pressure (2-3 mmH(2)O) halved the outflow of 24-40% O(2) venturi devices and increased oxygen concentration.
- The 60% O(2) venturi device delivered a maximum of 30 L/min at zero downstream pressure, with flow decreasing further under pressure.
- For all tested devices, increased downstream pressure led to decreased outflow and altered oxygen concentration, often falling below typical adult peak tidal flow.
Conclusions:
- Standard venturi devices are not consistently fixed performance under conditions of increased downstream resistance.
- Clinical use of venturi devices requires careful consideration of downstream resistance to ensure adequate oxygen delivery.
- Further research is needed to develop venturi systems with stable performance across varying physiological conditions.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
09:31Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
Related Concept Videos
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Steady Flow of a Fluid Stream
During this process, the momentum of the fluid within the control volume remains constant over the time interval dt. By applying the...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Bernoulli's Equation: Problem Solving
The first step is to compute the cross-sectional areas of the pipe and the Venturi throat to analyze the pressure difference indicated by the pressure gauge. Next, the continuity equation is...
Free Jet
Pipe Flowrate Measurement
The orifice meter is a simple,...