Related Experiment Video
Updated: Aug 23, 2026

Multicolor Fluorescence Detection for Droplet Microfluidics Using Optical Fibers
Published on: May 5, 2016
Temperature measurements of single droplets by use of laser-induced phosphorescence
Alaa Omrane1, Greger Juhlin, Frederik Ossler
1Division of Combustion Physics, Lund Institute of Technology, P.O. Box 118, Lund S-22100, Sweden. alaa.omrane@forbrf.lth.se
Abstract:
A novel technique for measuring droplet temperatures has been demonstrated. Laser-induced phosphorescence from thermographic phosphors, seeded to distilled water and iso-octane, was used to measure temperatures of single falling droplets. The phosphors were excited by the fourth and third harmonics of a Nd:YAG laser. The subsequent emission was evaluated by spectral and temporal investigations of the thermographic phosphors Mg4FGeO6:Mn and La2O2S:Eu, respectively. The spectral and the temporal methods permitted temperature measurements of free-falling droplets up to 433 K. Results from both methods, which show an estimated accuracy of better than 1%, are presented.
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
Photoluminescence: Applications
Flame Photometry: Lab