Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Interatrial defect sizing by intracardiac and transesophageal echocardiography compared with fluoroscopic
Fernando Boccalandro1, Andreas Muench, Joseph Salloum
1Division of Cardiology, University of Texas Medical School Houston, Houston, Texas, USA.
Abstract:
The purpose of this study was to evaluate prospectively the feasibility and accuracy of using echocardiographic measurements by transesophageal and intracardiac echocardiography (TEE and ICE, respectively) for interatrial septal defect sizing during percutaneous transcatheter closure. Forty-two patients underwent balloon sizing of interatrial septal defects using TEE in 21 of them and ICE in the other half. These measurements were correlated with quantitative fluoroscopic analysis and evaluated for bias and agreement between methods using a Bland-Altman analysis. Echocardiographic measurements were obtained by ICE and TEE in all patients. An excellent correlation was found between TEE and quantitative fluoroscopy (r = 0.898; P < 0.001) and between ICE and quantitative fluoroscopy (r = 0.876; P < 0.001), with a significant agreement (P < 0.001) and minimal positive bias toward the echocardiographic measurements. Both TEE and ICE are excellent methods of interatrial defect sizing when compared with quantitative fluoroscopic measurements.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022