Electrospray ionization with a pointed carbon fiber emitter
Jian Liu1, Kyung Won Ro, Mark Busman
1Department of Cell and Molecular Pharmacology and Experimental Therapeutics, Medical University of South Carolina, 173 Ashley Avenue, Charleston, SC 29425, USA.
Analytical Chemistry
|July 2, 2004
Summary
A novel carbon fiber electrospray ionization emitter enhances detection limits for mass spectrometry. This rugged device enables sensitive peptide analysis and interfacing with liquid chromatography-mass spectrometry systems.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Materials Science
Background:
- Mass spectrometry (MS) requires efficient ionization of analytes.
- Nanoliquid sampling techniques present challenges for interfacing with MS.
- Existing electrospray ionization (ESI) emitters have limitations in efficiency and robustness.
Purpose of the Study:
- To develop a new electrospray ionization emitter for improved nanoliquid sampling.
- To enhance detection limits in mass spectrometry.
- To create a robust and versatile interface for liquid chromatography-mass spectrometry.
Main Methods:
- Fabrication of a pointed carbon fiber emitter with a hydrophobic surface.
- Confining a small Taylor cone at the emitter tip for stable electrospray generation.
- Testing the emitter with standard peptides (angiotensin I) and complex peptide mixtures (cytochrome c and myoglobin tryptic digests).
- Evaluating the emitter's performance in interfacing with monolithic capillary HPLC.
Main Results:
- The carbon fiber emitter generated a stable electrospray with improved efficiency and enhanced detection limits.
- Ion current comparable to commercial nanospray emitters was achieved.
- Sensitive protein identification (<1 fmol peptide consumed) using ESI-MS/MS was demonstrated.
- Successful interfacing of monolithic capillary HPLC to MS was shown.
Conclusions:
- The pointed carbon fiber emitter offers a rugged and efficient solution for nanoliquid sampling in mass spectrometry.
- This technology significantly enhances sensitivity for peptide analysis.
- The emitter facilitates seamless integration of HPLC with MS for complex sample analysis.