Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
A beveled tip sheath liquid interface for capillary electrophoresis-electrospray ionization-mass spectrometry
Mei-Chun Tseng1, Yet-Ran Chen1, Guor-Rong Her1
1Department of Chemistry, National Taiwan University, Taipei, Taiwan, R.O.C.
Abstract:
A simple and durable sheath liquid interface for capillary zone electrophoresis-electrospray ionization-mass spectrometry (CZE-ESI-MS) has been developed. This interface utilized a beveled tip emitter and was found to be more sensitive than the conventional sheath liquid interface. The use of a beveled tip reduces the optimal flow rate and therefore decreases sample dilution. The interface utilized a 380 microm inner diameter and 400 microm outer diameter beveled tapered tip. Because of the large inner diameter and outer diameter of the tip, the interface is robust and can be easily implemented. The performance of this interface for CZE-ESI-MS and micelle electrokinetic capillary electrophoresis-electrospray-mass spectrometry, as demonstrated by the analysis of synthetic drugs and triazine mixtures, was significantly better than results obtained using a conventional sheath liquid interface.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...