Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Determination of nanogram level of VSCs using GC-SCD and an adsorption tube
1Department of Civil & Environmental Engineering, Sungkyunkwan University, Suwon, Korea. choiuk@skku.edu
Abstract:
A quantitative analytical method was used for detection of low level of volatile sulfur compounds (VSCs) using GC with a sulfur chemiluminescence detector (SCD) in this study. A linear response over the range of 2-90 ngS injected was obtained with a good repeatability or reproducibility. Equimolar response for H2S, DMS, MeSH, and EtSH was obtained by use of SCD and the response of the SCD is nearly equimolar for different sulfur compounds. It was possible to quantify the total VSCs as well as individual VSC using one of the standard VSCs. VSC recovery was measured with respect to some storing methods. An adsorption tube packed with molecular sieve 5A showed almost perfect recovery for both H2S and dimethyl sulfide (DMS) during 6 days at a dark state below 2 degrees. Whereas, with a gas tight pyrex vial or tedlar bag, it was impossible to obtain the recovery of 50% after 6 days. It is strongly recommended to use the adsorption tube for determination of nanogram levels of volatile sulfur compounds and for storing and concentrating VSCs effectively with a minimum experimental error.
More Related Videos
10:13Using Capillary Electrophoresis to Quantify Organic Acids from Plant Tissue: A Test Case Examining Coffea arabica Seeds
Published on: November 12, 2016
05:29Profiling Volatile Compounds in Blackcurrant Fruit using Headspace Solid-Phase Microextraction Coupled to Gas Chromatography-Mass Spectrometry
Published on: June 9, 2021
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...