Related Experiment Video
Updated: Aug 19, 2026

Sample Drift Correction Following 4D Confocal Time-lapse Imaging
Published on: April 12, 2014
A comparison of rotation- and blob-based system models for 3D SPECT with depth-dependent detector response
1Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI, USA. nastazia@eecs.umich.edu
Abstract:
We compare two different implementations of a 3D SPECT system model for iterative reconstruction, both of which compensate for non-uniform photon attenuation and depth-dependent system response. One implementation performs fast rotation of images represented using a basis of rectangular voxels, whereas the other represents images using a basis of rotationally symmetric volume elements. In our simulations the blob-based approach was found to slightly outperform the rotation-based one in terms of the bias-variance tradeoff in the reconstructed images. Their difference can be significant, however, in terms of computational load. The rotation-based method is faster for many typical SPECT reconstruction problems, but the blob-based one can be better-suited to cases where the reconstruction algorithm needs to process one volume element at a time.
Related Concept Videos
IR Spectroscopy: Molecular Vibration Overview
Stretching vibrations are vibrational motions that occur along the bond line, changing the bond length or distance between two bonded atoms. They are further distinguished as symmetric or asymmetric. In symmetric stretching, the...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...
IR Spectrometers