Related Experiment Video
Updated: Aug 23, 2026

Mass Spectrometry and Luminogenic-based Approaches to Characterize Phase I Metabolic Competency of In Vitro Cell Cultures
Published on: March 28, 2017
Rapid and sensitive liquid chromatography-tandem mass spectrometric method for the quantitation of metformin in human
Yingwu Wang1, Yunbiao Tang, Jingkai Gu
1Research Center for Drug Metabolism, Jilin University, Changchun 130023, China.
Abstract:
A rapid and sensitive liquid chromatography-tandem mass spectrometric (LC-MS-MS) method for the determination of metformin in human plasma using phenformin as internal standard has been developed and validated. Sample preparation of plasma involved acidification with acetic acid, deproteination with acetonitrile and washing with dichloromethane. Samples were then analyzed by HPLC on a short Nucleosil C18 column (5 microm, 50 mm x 4.6 mm i.d.) using a mobile phase consisting of acetonitrile:methanol:10mM ammonium acetate pH 7.0 (20:20:60, v/v/v) delivered at 0.65 ml/min. Detection was performed using an Applied Biosystems Sciex API 4000 mass spectrometer set at unit resolution in the multiple reaction monitoring (MRM) mode. Atmospheric pressure chemical ionization (APCI) was used for ion production. The assay was linear over the range 1-2000 ng/ml with intra- and inter-day precision of <8.6% and accuracy in the range 91-110%. The limit of detection was 250 pg/ml in plasma. The method was successfully applied to a clinical pharmacokinetic study of an extended-release tablet of metformin hydrochloride (500 mg) administered as a single oral dose.