A multiwire proportional counter for very high counting rates
A F Barbosa1, G P Guedes, E Tamura
1CBPF, Rua Dr Xavier Sigaud 150, CEP 22290-180, Rio de Janeiro, RJ, Brazil.
Abstract:
Preliminary measurements in a proportional counter with two independently counting wires showed that counting rates up to 10(6) counts s(-1) wire(-1) can be reached without critical loss in the 'true versus measured' linearity relation. Results obtained with a detector containing 30 active wires (2 mm pitch) are presented. With each wire is associated a fast pre-amplifier and a discriminator channel. Global counting rates in excess of 10(7) events s(-1) are reported. Dead-time losses are corrected by use of simple mathematical-modelling functions. Data-acquisition systems are described for one-dimensional (real-time) and two-dimensional (off-line) position-sensitive detection systems.