ASCANIO: a backscattering X-ray multichannel spectrometer for synchrotron applications
Pietro Barcella1,2, Giacomo Ticchi1,2, Beatrice Pedretti1,2
1Dipartimento di Elettronica, Informazione e Bioingegneria (DEIB), Politecnico di Milano, Milano, Italy.
Journal of Synchrotron Radiation
|August 5, 2026
Summary
ASCANIO, a new 16-channel X-ray spectrometer, enhances X-ray fluorescence mapping speed using silicon drift detectors. Its tilted design optimizes performance, achieving high count rates and excellent energy resolution for advanced synchrotron measurements.
Area of Science:
- Physics
- Materials Science
- Analytical Chemistry
Background:
- High count rate X-ray spectrometers are crucial for accelerating synchrotron measurements, particularly X-ray fluorescence mapping (XFM).
- Monolithic multichannel silicon drift detectors (SDDs) offer a path to developing dense, high-performance detector arrays for this purpose.
Purpose of the Study:
- To develop and characterize ASCANIO, a 16-channel backscattering X-ray spectrometer utilizing monolithic SDD modules.
- To optimize solid angle and count-rate uniformity through a novel 'tilted' configuration of SDD units.
Main Methods:
- Development of a 16-channel backscattering X-ray spectrometer (ASCANIO) based on monolithic silicon drift detector (SDD) modules.
- Implementation of a 'tilted' configuration for SDD units to enhance detection uniformity.
- Laboratory testing including cryogenic cooling and vacuum conditions, followed by spectroscopic performance verification.
- Commissioning on a synchrotron beamline (P06 at PETRA III) to evaluate performance under operational conditions.
Main Results:
- The ASCANIO system achieved a resolution of 134 eV FWHM at the Mn Kα peak.
- Under synchrotron operation, the spectrometer reached a cumulative output count rate of 20 Mcps with 20% dead time.
- The tilted configuration demonstrated positive effects on performance at the beamline.
Conclusions:
- ASCANIO represents a significant advancement in X-ray spectrometer technology for high-speed synchrotron applications.
- The developed spectrometer is suitable for demanding XFM analyses, as demonstrated by its application to a biological sample.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Scanning Electron Microscopy
A scanning electron microscope (SEM) is used to study the surface features of a sample by using an electron beam that scans the sample surface in a two-dimensional manner. Typically, areas between ~1 centimeter to 5 micrometers in width can be imaged. SEM can be used to image bacteria, viruses, tissues as well as larger samples like insects. Conventional SEM gives a magnification ranging from 20X to 30,000X and spatial resolution of 50 to 100 nanometers.
Fundamental Principles
Accelerated...
Fundamental Principles
Accelerated...
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...