Accuracy of the Tonosafe disposable tonometer head compared to the Goldmann tonometer alone
Peter Kim1, Somsak Lertsumitkul, Meaghan Clark
1Department of Ophthalmology, Liverpool Hospital, Sydney, New South Wales, Australia.
Purpose:
To assess the accuracy of the Tonosafe disposable prism tonometer head via a randomised controlled prospective trial.
Methods:
The intraocular pressure (IOP) of 69 patients (31 men, 38 women) involving 137 eyes was measured using both the Tonosafe disposable and the Goldmann tono-meter head. The average age of patients was 66.5 years (range 23-93 years). The two examiners (PM and SL) who performed tonometry were masked while a separate observer read the IOP measurements. The order was randomised between Tonosafe and Goldmann devices. The minimum interval between the two measurements was 15 min. Patients with corneal and external diseases were excluded from the study.
Results:
The mean IOP using the Goldmann tonometer head was 17.44 +/- 4.97 mmHg. The mean IOP using the Tonosafe disposable head was 17.58 +/- 5.03 mmHg. The mean difference was 0.14 +/- 1.73 mmHg (95% CI -0.44, 0.16, P= 0.36). Subgroup analysis on eyes with IOP > 21 mmHg showed the Tonosafe disposable head was on average 0.15 +/- 2.40 mmHg higher than the Goldmann tonometer head. There was a high level of correlation between the IOPs obtained with the Goldmann and Tonosafe heads. Pearson's coefficient of correlation was 0.94 (P < 0.0001).
Conclusion:
The Tonosafe disposable prism head was found to be accurate in IOP measurement, even in the higher range.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Comparing Experimental Results: Student's t-Test