In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
[Echographic calculation of left ventricular mass in TM mode]
F Martin1, N Mansencal, E Abergel
1Service de cardiologie, Assistance publique-Hôpitaux de Paris, hôpital Ambroise Paré, Boulogne.
Summary
New imaging techniques like harmonic (H) and colour tissue Doppler (CTD) overestimate left ventricular mass (LVM) compared to fundamental (F) imaging. Harmonic imaging offers better reproducibility for LVM measurements.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Transthoracic (TM) measurements of left ventricular mass (LVM) are typically validated using fundamental (F) imaging.
- Newer imaging modalities, harmonic (H) and colour tissue Doppler (CTD), may impact TM measurements of LVM.
Purpose of the Study:
- To evaluate the effect of harmonic (H) and colour tissue Doppler (CTD) imaging on transthoracic (TM) measurements of left ventricular mass (LVM).
- To compare these measurements against traditional fundamental (F) imaging and assess inter-observer reproducibility.
Main Methods:
- A prospective study involving 26 patients with valid TM sections.
- Left ventricular (LV) and LVM parameters were assessed using F, H, and CTD modes, applying both Penn and ASE conventions.
- Inter-observer reproducibility was analyzed for all imaging modes.
Main Results:
- High correlations were observed between F and H (r>0.95) and F and CTD (r>0.95) for LVM measurements, irrespective of convention.
- LVM measurements in H and CTD modes were consistently higher than in F mode (p<0.02).
- False diagnoses of LV hypertrophy occurred in 27% of patients with H and 15% with CTD. Harmonic imaging demonstrated superior inter-observer reproducibility (23+/-15) compared to F (32+/-19) and CTD (59+/-18).
Conclusions:
- Harmonic (H) and CTD imaging lead to an overestimation of LVM, primarily due to increased parietal thickness measurements.
- Inter-observer reproducibility was excellent for H imaging but poor for CTD.
- New formulae are required for validated LVM calculations using harmonic imaging.