Hold-and-flush, a novel fraction collection method in semi-preparative subcritical and supercritical fluid
Jiejun Wu1, Heather McAllister
1Drug Discovery, Johnson and Johnson Pharmaceutical Research and Development, LLC, 3210 Merryfield Row, San Diego, CA 92121-1126, USA. jwu7@prdus.jnj.com
Abstract:
In supercritical fluid chromatography, the partial phase separation that occurs in the dead volume between automatic backpressure regulator and collection vessels causes significant peak tailing and delayed arrival of compounds to the collectors. As a result, when two peaks are barely baseline separated, it becomes very difficult to correctly set fraction collection triggers, which in turn results in fractions being collected with lower purity and lower yield. The problem can be solved with a simple addition of a four-port, two-position switching valve between UV detector and automatic back pressure regulator. The valve acts as a timed gate to release each peak out of the back pressure regulator and into the collection vessel, while at the same time it holds later peaks inside the UV detector and the column in a closed loop.
Related Concept Videos
Principles Of Column Chromatography
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...