Quantitative Hardness Measurement by Instrumented AFM-indentation
Published on: November 22, 2016
Study of hardness and roughness modification in explanted joint components
L Torrisi1, A M Visco, N Campo
1Physics Department, University of Messina, Salita Sperone 31, I-98166 Messina, Italy. lorenzo.torrisi@unime.it
Abstract:
The significant wear of the UHMWPE bearings of explanted knee prostheses is produced mainly by micrometric debris ("third-body" wear) that diffuse toward the mobile metal-polymer interface. Here debris is crushed during the movement producing scratches in the metal and in the polymeric surfaces. Mechanical stress and the biological effects change the physical polymeric properties. In order to evaluate the area of UHMWE bearings submitted to high load stresses, in this work physical investigations are performed on the explanted knee prosthesis. Particularly, the roughness profile analysis (RPA) and the micro-hardness measurements (MHM) resulted suitable for the localisation of the mechanical and biological wear area. In the stressed zone, surface treatments could be applied in order to improve the mechanical resistance of the polymeric material. Particularly, the ion implantation with heavy ions is proposed to enhance the polymeric wear resistance. The Scanning Electron Microscopy (SEM), Energy Dispersive X-ray Spectrometry (EDX) and Infrared absorption spectroscopy (FTIR) techniques were also applied and results discussed.
Related Concept Videos
Toughness and Hardness of Aggregate
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used to...