Prostate volume ultrasonography: the influence of transabdominal versus transrectal approach, device type and
J W N C Huang Foen Chung1, S H de Vries, R Raaijmakers
1Department of Urology, Sector Furore, Erasmus MC, Room EE 1630, PO Box 1738, 3000 DR Rotterdam, The Netherlands.
Objectives:
We conduct a longitudinal non-invasive study of changes in urinary bladder contractility secondary to benign prostatic enlargement. In that study, the prostate volume is estimated by transabdominal ultrasonography. The accuracy of those measurements was verified by comparison of transabdominal to transrectal stepwise planimetric ultrasonography as the gold standard. Also, two different transabdominal devices used were compared, and the influence of different operators was studied.
Materials & Methods:
Two series of measurements in 100 patients each were done. In the first series, transabdominal and transrectal sonography were pairwise compared in each patient. In the second series, transabdominal measurements were done with two devices (a hospital Aloka SSD-1700 and a portable Aloka SSD-900). Transrectal scannings were done by three investigators whilst all transabdominal scannings were done by one. Regression graphs, ratio plots and statistical analyses of the data quantified the reproducibility of different methods, observers and device types.
Results:
In the transrectal-transabdominal series of prostate volume measurements (in cm3), the Pearson correlation coefficient was 0.84 (p < 0.001), the mean of the means was 51.8 +/- 23.0 (mean +/- S.D.), and the mean of the differences was 1.0 +/- 1.4. In the series with two devices, the Pearson correlation coefficient was 0.73 (p < 0.001), the mean of the means was 31.0 +/- 10.9, and the mean of the differences was 1.0 +/- 1.3.
Conclusion:
No statistically significant differences were found between the transabdominal- transrectal ultrasonography, two different transabdominal devices nor between different observers. However, for those using these measurements in everyday clinical practice, it is worth to point out that in our data a transabdominal scan and a transrectal scan in the same patient, on the same day, differed more than 30% in one fourth of the patients and that two transabdominal scans in the same patient (with two different devices, on two different days) differed more than 30% in every fifth patient.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies II: Ultrasonography