Radiation doses of indirect and direct digital cephalometric radiography
F Gijbels1, G Sanderink, J Wyatt
1Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Katholieke Universiteit Leuven, Belgium.
Aim:
The aim of this study was to measure organ doses and calculate the effective dose for indirect and direct digital cephalometric exposures.
Material And Methods:
Indirect digital cephalometric exposures were made of a Rando phantom head using a Cranex Tome multipurpose unit with storage phosphor plates from Agfa and the direct digital (Charge Coupled Device, CCD) exposures were made with a Proline Ceph CM unit. Exposure settings were 70 kV and 4 mAs for indirect digital exposures. Direct digital exposures were made with 70 kV, 10 mA and a total scanning time of 23 s. TLD700 dosemeters were used to measure organ doses, and the effective doses were calculated with (effective dose(sal)) and without inclusion of the salivary glands. A pilot study was carried out to compare diagnostic image quality of both imaging modalities.
Results:
Effective doses were 1.7 microSv for direct digital and 1.6 microSv for indirect digital cephalometric imaging. When salivary glands were included in the calculation, effective doses(sal) were 3.4 microSv and 2.2 microSv respectively. Organ doses were higher for direct digital imaging, except for the thyroid gland, where the organ doses were comparable. Diagnostic image quality of indirect and direct digital cephalometric images seemed comparable.
Conclusion:
Effective dose and effective dose(sal) were higher for direct digital cephalometric exposure compared with indirect digital exposure. Organ doses were higher for direct digital cephalography. From preliminary data, it may be presumed that diagnostic image quality of indirect and direct digital cephalometric images are comparable.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...