Related Experiment Video
Updated: Aug 22, 2026

A Straightforward Method for Glucosinolate Extraction and Analysis with High-pressure Liquid Chromatography (HPLC)
Published on: March 15, 2017
High-performance thin-layer chromatographic-densitometric determination of secoisolariciresinol diglucoside in
Silvia A Coran1, Valerio Giannellini, Massimo Bambagiotti-Alberti
1Dipartimento di Scienze Farmaceutiche, Università di Firenze, Via Ugo Schiff 6, I-50019 Sesto Fiorentino (Florence), Italy. coran@unifi.it
Abstract:
A HPTLC-densitometric method, based on an external standard approach, was developed in order to obtain a novel procedure for routine analysis of secoisolariciresinol diglucoside (SDG) in flaxseed with a minimum of sample pre-treatment. Optimization of TLC conditions for the densitometric scanning was reached by eluting HPTLC silica gel plates in a horizontal developing chamber. Quantitation of SDG was performed in single beam reflectance mode by using a computer-controlled densitometric scanner and applying a five-point calibration in the 1.00-10.00 microg/spot range. As no sample preparation was required, the proposed HPTLC-densitometric procedure demonstrated to be reliable, yet using an external standard approach. The proposed method is precise, reproducible and accurate and can be employed profitably in place of HPLC for the determination of SDG in complex matrices.
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...