Proposed definitions for isodose flatness and symmetry in clinical radiotherapy beams
Eduardo Galiano1, Tara Joly, Federico Wiebe
1Department of Physics, Laurentian University, Ramsey Lake Road, Sudbury, Ont. P3E 2C6, Canada. egalianoriveros@laurentian.ca
Abstract:
In clinical radiotherapy it is important that beam intensity be as homogeneous as possible to reduce the probability of treatment failure. As an extension of the well established concepts of beam flatness and symmetry for characterizing radiotherapy beams, the concepts of isodose flatness and symmetry are introduced. The definitions are tested with actual data obtained from a Co-60 unit and a linear accelerator and results are presented. The concepts as defined appear to provide a useful quantitative indication of beam homogeneity.