The radiation field characteristics of a 137Cs source used for calibration of radiation protection instruments
Related Experiment Video
Updated: Aug 22, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
E Fernandes1, D Freire, A C de Freitas
1Laboratório de Ciências Radiológicas/ DBB/IBRAG da Universidade do Estado do Rio de Janeiro, Rio de Janeiro, RJ-Rua São Francisco Xavier, 524. CEP: 20550-013. Rio de Janeiro. RJ Brazil.
Abstract:
In this work, an experimental study of the radiation beam characteristics of the photon beams produced by commercial irradiator that uses a single collimated 137Cs source is performed. A set of lead attenuators are placed at the exit window of the irradiator to vary the air Kerma that is required to cover the instrument scales at a particular calibration distance. The results strongly enhance the findings of previous studies that a single source with appropriate attenuation and collimation is a practical solution for calibration laboratories with an acceptable overall accuracy.