Efficient high-frequency body coil for high-field MRI.

J T Vaughan1, G Adriany, C J Snyder

  • 1Center for MR Research, Department of Radiology, School of Medicine, University of Minnesota, Minneapolis, Minnesota 55455, USA. tommy@cmrr.umn.edu

Magnetic Resonance in Medicine
|September 25, 2004
PubMed
Summary

Researchers developed a new transverse electromagnetic (TEM) body coil for MRI, enabling efficient imaging at 4 Tesla (T) and showing potential up to 8 T. This innovation addresses challenges in high-field MRI coil design.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Retrospective evaluation of canine and feline maxillomandibular trauma cases. A comparison of signalment with non-maxillomandibular traumatic injuries (2003-2012).

Veterinary and comparative orthopaedics and traumatology : V.C.O.T·2014
Same author

Retrospective evaluation of canine and feline maxillomandibular trauma cases. Comparison of lunar cycle and seasonality with non-maxillomandibular traumatic injuries (2003-2012).

Veterinary and comparative orthopaedics and traumatology : V.C.O.T·2014
Same author

Prevalence and nature of dentoalveolar injuries among patients with maxillofacial fractures.

The Journal of small animal practice·2012
Same author

Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T.

Magnetic resonance in medicine·2011
Same author

7 Tesla (T) human cardiovascular magnetic resonance imaging using FLASH and SSFP to assess cardiac function: validation against 1.5 T and 3 T.

NMR in biomedicine·2011
Same author

Initial results of cardiac imaging at 7 Tesla.

Magnetic resonance in medicine·2008

Area of Science:

  • Magnetic Resonance Imaging (MRI)
  • Radiofrequency (RF) Coil Engineering
  • Medical Physics

Background:

  • Body coils are crucial for homogeneous excitation in MRI, often combined with surface coils for reception.
  • Designing efficient body coils for high field strengths (3T and above) presents significant engineering challenges due to physical size and electrical length.
  • Existing body coil designs struggle to meet the demands of higher field strengths, limiting advanced MRI applications.

Purpose of the Study:

  • To develop an efficient transverse electromagnetic (TEM) body coil suitable for high-field MRI applications.
  • To evaluate the performance of the new TEM body coil in human studies at field strengths up to 4 Tesla (T).
  • To assess the feasibility of the developed body coil technology for even higher field strengths, such as 8 T.

Related Experiment Videos

Main Methods:

  • Development of a novel transverse electromagnetic (TEM) body coil design.
  • Implementation and testing of the TEM body coil in human MRI studies up to 4 T.
  • Acquisition of head, body, and breast images within specified peak power constraints (<8 kW).
  • Bench studies to determine the coil's feasibility at higher field strengths (up to 8 T).
  • Application of radiofrequency (RF) shimming to mitigate high-field related artifacts, specifically in cardiac imaging.

Main Results:

  • The developed TEM body coil successfully enabled human imaging at field strengths up to 4 T.
  • High-quality head, body, and breast images were acquired within acceptable peak power limits.
  • Bench studies suggest the coil design is viable for operation at field strengths up to 8 T.
  • RF shimming proved effective in resolving a cardiac imaging artifact observed at high fields.

Conclusions:

  • The novel TEM body coil represents a significant advancement for high-field MRI, overcoming previous design limitations.
  • This technology facilitates efficient and safe human imaging at 4 T and shows promise for future ultra-high-field MRI systems (8 T).
  • The developed coil and RF shimming technique contribute to improved image quality and artifact reduction in advanced MRI.