Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T
C J Snyder1, L Delabarre, S Moeller
1Department of Radiology, University of Minnesota, Minneapolis, MN, USA.
Abstract:
Eight- and sixteen-channel transceive stripline/TEM body arrays were compared at 7 T (297 MHz) both in simulation and experiment. Despite previous demonstrations of similar arrays for use in body applications, a quantitative comparison of the two configurations has not been undertaken to date. Results were obtained on a male pelvis for assessing transmit, signal to noise ratio, and parallel imaging performance and to evaluate local power deposition versus transmit B(1) (B(1) (+) ). All measurements and simulations were conducted after performing local B(1) (+) phase shimming in the region of the prostate. Despite the additional challenges of decoupling immediately adjacent coils, the sixteen-channel array demonstrated improved or nearly equivalent performance to the eight-channel array based on the evaluation criteria. Experimentally, transmit performance and signal to noise ratio were 22% higher for the sixteen-channel array while significantly increased reduction factors were achievable in the left-right direction for parallel imaging. Finite difference time domain simulations demonstrated similar results with respect to transmit and parallel imaging performance, however, a higher transmit efficiency advantage of 33% was predicted. Simulations at both 3 and 7 T verified the expected parallel imaging improvements with increasing field strength and showed that, for a specific B(1) (+) shimming strategy used, the sixteen-channel array exhibited lower local and global specific absorption rate for a given B(1) (+) .
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies IV: Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...