Performing daily prostate targeting with a standard V-EPID and an automated radio-opaque marker detection algorithm
Luc Beaulieu1, Louis-Martin Girouard, Sylviane Aubin
1Department of radiation oncology and Centre de Recherche en Cancérologie de l'Université Laval, Hôtel-Dieu de Québec, Centre Hospitalier Universitaire de Québec (CHUQ), 11 Cote du Palais, Québec, Canada, G1R 2J6.
Abstract:
Online prostate positioning using gold markers and a standard video-based electronic portal imaging device is reported. The average systematic (random) errors have been reduced from 2.1 mm (2.7 mm) to 0.5 mm (1.5 mm) in AP direction, 1.1 mm (1.7 mm) to 0.7 mm (1.2 mm) SI and 1.2 mm (1.7 mm) to 0.6 mm (1.3 mm) LR.