The borderline or weakly positive Hybrid Capture II HPV test: a statistical and comparative (PCR) analysis
Jerome M Federschneider1, Liping Yuan, Joshua Brodsky
1Department of Obstetrics and Gynecology, Brigham and Women's Hospital, Boston, Mass 02115, USA.
Objectives:
Recent studies have hypothesized that laboratory contamination may influence interpretation of Hybrid Capture II (HCII) human papillomavirus (HPV) detection assay values.
Study Design:
To test this hypothesis, 572 consecutive HCII samples were statistically evaluated to test the null hypothesis that cross-well contamination was not present. In addition, 874 consecutive paired samples from patients followed by both HCII and polymerase chain reaction (PCR) analysis were compared.
Results:
A Kendall's tau measure of association among adjacent wells yielded a P value of .016, rejecting the null hypothesis of no contamination. Analysis of relative light unit values between 0.8 and 1.5 rejected the null hypothesis at P=.077. Moreover, PCR positivity was significantly higher for samples with HCII values above 1.5 vs 0.8 to 1.5 (P=.001).
Conclusion:
Cross-well contamination of samples occurs during processing, and may influence interpretation of some borderline positives. The proportion of cases at risk is low (<3%). Nevertheless, this information may be germane to the interpretation and reporting of marginally positive HCII test values.