Rationale And Objectives:
Metallic guide wires can be subject to substantial heating when used in the magnetic resonance (MR) environment. Therefore, animal experiments were performed to test the feasibility of a non-metallic and MR-safe guide wire with passive markers for catheterization of coronary arteries under MR guidance.
Materials And Methods:
Self-made guide wires consisting of a resin-microparticle compound covered by polytetrafluoroethylene were used to catheterize both coronary arteries of swine together with a non-braided catheter. Time needed for catheterization was recorded.
Results:
MR-guided coronary artery catheterization with passive visualization of a self-made non-metallic guide wire is possible. In average 141 seconds (SD 68) were needed to manipulate the guide wire together with a catheter from the carotid artery into the left or right coronary artery ostium.
Conclusion:
Standard nitinol guide wires have to be considered unsafe for MR-guided interventions due to possible heating of electrical conducting structures in the MR environment. Passive visualization techniques allow MR-guided catheterization of small arteries like coronaries. However, there is the substantial disadvantage of obscuring the underlying anatomy of small vessels by the passive markers needed for real-time MR guidance.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging