Heuristic evaluation of infusion pumps: implications for patient safety in Intensive Care Units
Mark J Graham1, Tate K Kubose, Desmond Jordan
1Department of Biomedical Informatics, College of Physicians and Surgeons, Columbia University, USA.
Objective:
The goal of this research was to use a heuristic evaluation methodology to uncover design and interface deficiencies of infusion pumps that are currently in use in Intensive Care Units (ICUs). Because these infusion systems cannot be readily replaced due to lease agreements and large-scale institutional purchasing procedures, we argue that it is essential to systematically identify the existing usability problems so that the possible causes of errors can be better understood, passed on to the end-users (e.g., critical care nurses), and used to make policy recommendations.
Design:
Four raters conducted the heuristic evaluation of the three-channel infusion pump interface. Three raters had a cognitive science background as well as experience with the heuristic evaluation methodology. The fourth rater was a veteran critical care nurse who had extensive experience operating the pumps. The usability experts and the domain expert independently evaluated the user interface and physical design of the infusion pump and generated a list of heuristic violations based upon a set of 14 heuristics developed in previous research. The lists were compiled and then rated on the severity of the violation.
Results:
From 14 usability heuristics considered in this evaluation of the Infusion Pump, there were 231 violations. Two heuristics, "Consistency" and "Language", were found to have the most violations. The one with fewest violations was "Document". While some heuristic evaluation categories had more violations than others, the most severe ones were not confined to one type. The Primary interface location (e.g., where loading the pump, changing doses, and confirming drug settings takes place) had the most occurrences of heuristic violations.
Conclusion:
We believe that the Heuristic Evaluation methodology provides a simple and cost-effective approach to discovering medical device deficiencies that affect a patient's general well being. While this methodology provides information for the infusion pump designs of the future, it also identifies important insights concerning equipment that is currently in use in critical care environments.
More Related Videos
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Heart Failure VI: Adjunct Therapies
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...