Related Experiment Video
Updated: Aug 21, 2026

Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Half-life determination of 88Y and 89Sr
M N Amiot1, J Bouchard, M M Bé
1BNM-CEA/LNHB, Laboratoire National Henri Becquerel, 91191 Gif-sur-Yvette, Cedex, 31062 France. marie-noelle.amiot@cea.fr
Abstract:
Half-life measurements have been carried out at LNHB for 88Y using a 4pi gamma-ionisation chamber and for 89Sr using a proportional counter. The determined half-life values and associated standard uncertainties are 106.63 +/- 0.05 d for 88Y and 50.65 +/- 0.05 d for 89Sr, being consistent with relevant values reported in literature. Based on the present results and relevant literature values revised recommended half-life values and associated standard uncertainties are proposed, viz. 106.626 +/- 0.021 d for 88Y and 50.57 +/- 0.03 d for 89Sr.
Related Concept Videos
Radioactive Decay and Radiometric Dating
Nuclear Stability
To hold positively charged protons together in the...
Half-life of a Reaction
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Isotopes and Radioisotopes
An isotope containing more...
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay: