Safety and feasibility of using a central venous catheter for rapid contrast injection rates
Pina C Sanelli1, Monica Deshmukh, Igor Ougorets
1Department of Radiology, New York Presbyterian Hospital, Weill Medical College of Cornell University, 520 E 70th St., Starr Pavilion-630, New York, NY 10021, USA. pcs9001@med.cornell.edu
Objective:
Our aim was to determine the safety and feasibility of using a central venous catheter for rapid contrast injections during CT.
Materials And Methods:
An in vitro experiment was performed using a 7-French Arrow-Howes multilumen central venous catheter. Each catheter port was tested by varying contrast agent flow rates delivered by a power injector. Contrast media specifications were kept similar to routine clinical practice. The in vivo experiment included 104 cases in which rapid contrast injections, 3.0-5.0 mL/sec, were delivered through a central venous catheter for dynamic CT examinations. Patient monitoring for early complications of contrast extravasation, cardiac arrhythmia, and allergic reactions was performed. Contrast injections were monitored for pressure limitation, automatic flow-rate adjustment, and catheter injury. Chart review was performed for delayed complications of mediastinal hematoma, infection, or catheter malfunction.
Results:
During the in vitro experiment, all desired flow rates, 3.0-9.9 mL/sec, could be delivered through the central venous catheter with no catheter injury. No immediate or early patient or catheter complications were observed during the in vivo experiment. Follow-up evaluation revealed that 18 blood cultures and one catheter culture were positive for bacterial growth. In a subgroup of 43 patients, five contrast injections were pressure-limited by the power injector, and only one had the flow rate automatically adjusted to 3.6 mL/sec from 4.0 mL/sec.
Conclusion:
Rapid contrast injection rates, at 3.0-5.0 mL/sec, through the Arrow-Howes multilumen central venous catheter are feasible and safe in the clinical setting. However, a strict protocol should be followed to avoid possible serious complications.
More Related Videos
Related Concept Videos
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging