M C Boqué1, B Gualis, A Sandiumenge
1University Hospital Joan XXIII, University Rovira i Virgili, Mallafré Guasch 4, 43007 Tarragona, Spain. cboque@hjxxiii.scs.es
Objective:
To quantify the incidence and degree of endotracheal tube intraluminal obstruction after mechanical ventilation and its relation to time of intubation.
Design:
Prospective observational study.
Setting:
A 14-bed medical-surgical intensive care unit at a university-affiliated teaching hospital.
Patients:
Ninety-four endotracheal tubes used in 80 patients requiring mechanical ventilation for more than 12 h.
Interventions And Results:
Acoustic reflectometry was performed in every endotracheal tube after patient extubation to measure its volume reduction. The intraluminal volumes of used endotracheal tubes in mechanically ventilated patients were significantly lower than those of unused tubes of the same size (5.52+/-0.92 ml(3) versus 6.54+/-0.79 ml(3), p<0.05). The mean difference in endotracheal tube segment volumes was 15.2% (range 0-66%). Volume reduction was above 10% in 60.8% of the tubes. In 22% of endotracheal tubes the remaining inner diameter was less than 7 mm. Reduction below this figure was less frequent (9.3%) in tubes 8 mm or more (p<0.05). The percentage of endotracheal tube volume reduction was not associated with the duration of intubation (r=-0.09, p= n.s.) Peak pressure measured before extubation did not predict obstruction (r=0.11, p= n.s.).
Conclusions:
Inadvertent endotracheal tube obstruction was common in patients requiring mechanical ventilation and may be significant as early as at 24 h. Moderate obstruction in endotracheal tube lumens should be suspected in cases of difficulties in weaning, even in patients who were ventilated for less than 1 day.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...