Related Experiment Video
Updated: Aug 20, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Performance improvement of small gamma camera using NaI(Tl) plate and position sensitive photo-multiplier tubes
Myung Hwan Jeong1, Yong Choi, Yong Hyun Chung
1Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Kangnam-Ku, Seoul, 135-710, Korea.
Abstract:
The purpose of this study was to improve the performance of a small gamma camera, utilizing a NaI(Tl) plate and a 5" position sensitive PMT. We attempted to build a NaI(Tl) plate crystal system which retained all its advantages, while at the same time integrating some of the advantages inherent in an array-type scintillation crystal system. Flood images were obtained with a lead hole mask, and position mapping was performed by detecting hole positions in the flood image. Energy calibration was performed using the energy spectra obtained from each hole position. Flood correction was performed using a uniformity correction table containing the relative efficiency of each image element. The spatial resolution was improved about 16% after correction at the centre field of view. Resolution deterioration at the outer field of view (OFOV) was considerably ameliorated, from 6.7 mm to 3.2 mm after correction. The sensitivity at the OFOV was also increased after correction, from 0.7 cps microCi(-1) to 2.0 cps microCi(-1). The correction also improved uniformity, from 5.2% to 2.1%, and linearity, from 0.5 mm to 0 mm. The results of this study indicate that the revised correction method can be employed to considerably improve the performance of a small gamma camera using a NaI(Tl) plate-type crystal. This method also provides high spatial resolution and linearity, like array-type crystals do, while retaining the specific advantages of plate-type crystals.
More Related Videos
10:48Probing Nicotinic Acetylcholine Receptor Function in Mouse Brain Slices via Laser Flash Photolysis of Photoactivatable Nicotine
Published on: January 25, 2019
16:10A TIRF Microscopy Technique for Real-time, Simultaneous Imaging of the TCR and its Associated Signaling Proteins