Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Increased digitalis-like immunoreactive substances in neonatal plasma measured using fluorescence polarization
Objective:
To better define the reported increased digitalis-like immunoreactive substances (DLIS) in neonatal plasma, we studied the relation among plasma DLIS level, blank intensity (BLK-I) value at FPIA measurement and plasma total bilirubin level.
Methods:
The DLIS levels were measured in 10 neonates with or without jaundice and 10 infants in good health, using fluorescence polarization immunoassay (FPIA) and microparticle enzyme immunoassay (MEIA). BLK-I value and plasma total bilirubin level were also measured simultaneously.
Results:
In neonates with jaundice, DLIS using FPIA, BLK-I and total bilirubin level were 0.58 +/-0.13 ng/mL, 2598 +/- 408, and 17.98 +/- 1.13 mg/dL, respectively, before phototherapy, and 0.33 +/-0.06 ng/mL, 1886 +/- 237, and 15.16 +/- 2.07 mg/dL after phototherapy. Corresponding values in neonates without jaundice were (DLIS: 0.34 +/-0.04 ng/mL; BLK-I: 1,764 +/- 278; total bilirubin: 10.37 +/- 4.54 mg/dL); in healthy infants (0.12 +/-0.06 ng/mL, 400.7 +/- 4.6 and 0.42 +/- 0.13 mg/dL, respectively) and in healthy volunteers (0.10 +/-0.07 ng/mL, 403.1 +/- 8.4, and 0.58 +/- 0.30 mg/dL, respectively). Using MEIA, DLIS was not detected in 10 neonates, 10 infants and 20 healthy volunteers.
Conclusions:
A fluorescent compound related to bilirubin increased the BLK-I value in the measurement of neonatal plasma using FPIA. The fluorescence was not the result of endogenous digitalis-like factors.