Effect of a compensated Jaffe creatinine method on the estimation of glomerular filtration rate
Michael H M Chan1, K F Ng, C C Szeto
1Department of Chemical Pathology, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, Hong Kong, China.
Background:
Roche Diagnostics has issued new c-fas calibrators for its automated systems. These produce creatinine values that are more comparable with those obtained by high-performance liquid chromatography. However, this results in an underestimation of measured creatinine at concentrations below 155 micromol/L and an overestimation at concentrations above this value.
Methods:
Serum and urine creatinine concentrations were prospectively determined on samples from 60 patients using the new (compensated) and old (uncompensated) c-fas calibrators, and Passing-Bablok regression analysis was performed. The regression equations thus determined were then used retrospectively to determine the compensated creatinine results (i.e. those results that would have been obtained using the new calibrator) in those serum and urine samples analysed in the previous year using the old uncompensated c-fas calibrator. The compensated creatinine results were then used to estimate the glomerular filtration rate (GFR) by calculating creatinine clearance. This was done by using the formula: UV/Pt, in which U represents the urinary creatinine concentration (micromol/L), V the urinary collection volume (mL), P the serum creatinine concentration (micromol/L) and t the urinary collection time (min). It was also calculated using the abbreviated Modification of Diet in Renal Disease study group (MDRD) formula.
Results:
The creatinine clearance as determined using either the UV/Pt calculation or the MDRD formula overestimated GFR by approximately 30% and approximately 50%, respectively, in normal individuals with a serum creatinine concentration below 155 micromol/L. However, in patients with mild to moderate renal failure (serum creatinine from 155 to 500 micromol/L), changes in creatinine clearances determined by the two procedures were minimal.
Conclusion:
When laboratories introduce this new, compensated calibrator into practice, it may be appropriate to discuss its potential impact with clinical staff who monitor patients using creatinine clearance.
Related Concept Videos
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Renal Drug Excretion: Glomerular Filtration
Drugs gain access to the kidney via the renal artery, which progressively branches off into afferent arterioles.
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...