Related Experiment Video
Updated: Aug 20, 2026

Applications of Liquid-Chromatography Tandem Mass Spectrometry in Natural Products Research: Tropane Alkaloids as a Case Study
Published on: March 8, 2024
Determination of atractyloside in Callilepis laureola using solid-phase extraction and liquid
Paul A Steenkamp1, Nial M Harding, Fanie R van Heerden
1Forensic Chemistry Laboratory, Department of Health, Johannesburg 2000, South Africa. fcljhb@icon.co.za
Abstract:
A selective analytical method based on high-performance liquid chromatography, combined with atmospheric pressure ionisation mass spectrometry, was developed for the detection of atractyloside. The analysis was performed on an Xterra Phenyl column utilising a gradient elution profile and a mobile phase consisting of 10 mM aqueous ammonium acetate buffer-methanol-acetonitrile. The calibration curve of the method (1 ng/ml-160 microg/ml) was best described by a second order polynomial function (r2 = 0.998) but displayed good linearity in the range of 100 ng/ml-1 microg/ml (r2 = 0.999). The limit of detection for the atractyloside standard was determined and found to be 100 pg/ml and the limit of quantification of atractyloside in tuber matrix was found to be 250 pg/ml. The relative standard deviation of the method was on average below 5% (n = 8). The method was successfully applied to the analysis of Callilepis laureola tubers and unknown powdered samples for the presence of atractyloside.