Detectability of catheters on bedside chest radiographs: comparison between liquid crystal display and
Martina Scharitzer1, Mathias Prokop, Michael Weber
1Department of Radiology and Ludwig Boltzmann-Institute for Clinical and Experimental Radiologic Research, University of Vienna, Währinger Gürtel 18-20, A-1090 Vienna, Austria. martina.scharitzer@meduniwien.ac.at
Purpose:
To compare observer performance with a flat-panel liquid crystal display (LCD) monitor and with a high-resolution gray-scale cathode-ray tube (CRT) monitor in the detection of simulated support catheters on bedside chest radiographs.
Materials And Methods:
The ethics committee did not require approval or patient informed consent when this study began. Because of a change in regulations, before images were acquired the nature of the study and procedures were explained to patients or their relatives, and consent was then obtained. A total of 131 catheter fragments (12-14 per radiograph) were superimposed over 10 anteroposterior bedside chest radiographs obtained with storage phosphor technology. Images were displayed on an LCD monitor (1536 x 2048 matrix) and a CRT monitor (2048 x 2560 matrix). Five radiologists independently located the catheter fragments and rated their confidence in detection with bright and subdued ambient light. A two-way analysis of variance and the Friedman test were used for statistical analysis.
Results:
There was no significant difference for either display type with respect to correctly detected catheter fragments (mean sensitivity, 56.6% and 56.0% for the CRT and the LCD monitors, respectively, with bright light and 61.2% for both monitors with subdued light). With both display types, detection rate with bright light decreased significantly (P < .05). False-positive rates and confidence ratings were not significantly affected by monitor type or ambient light.
Conclusion:
In a study with simulation of clinical conditions, performance of the LCD monitor and high-resolution CRT monitor for detection of support catheters on bedside chest radiographs was equivalent. With both displays, detection performance was equally reduced with bright ambient light.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Holter Monitor: 24-Hour Monitoring