Related Experiment Video
Updated: Aug 20, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Dosimetric consequences of manual pullback procedure for coronary artery radiotherapy with 90Sr/90Y beta-source
Ileana Iftimia1, Phillip M Devlin, Andre-Guy Martin
1Department of Radiation Oncology, Brigham and Women's Hospital, Boston, MA, USA. ileana.n.iftimia@lahey.org
Purpose:
This work presents a quantitative dosimetric analysis of the Novoste (90)Sr/(90)Y beta-source cardiovascular brachytherapy treatments using a manual pullback technique for patients with in-stent restenosis.
Methods And Materials:
Based on our previous measurements, a model was developed to estimate the dose in the middle of the junction region for tandem irradiation expressed as fraction of prescription dose (FPD) and dosimetric overlap length (DOL) receiving more/less than a threshold dose. The overlap/gap size was measured using the digital cine images recorded during treatment and then FPD and DOL were quantified.
Results:
Statistical analysis of 55 patients showed that the overlap size and the FPD at 2 mm radial distance were in range of 0 to 23 mm and 13-200% of prescription dose (Rx), respectively. Four gaps out of 76 pullback cases were found, but their size was at most 5 mm.
Conclusion:
Use of a 5 mm overlap avoided underdosed regions in the vast majority of the cases. These results are the first step towards an analysis of the clinical outcome of these patients.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...