Band broadening in size-exclusion chromatography of polydisperse samples
Simona-Tereza Popovici1, Wim Th Kok, Peter J Schoenmakers
1Polymer-Analysis Group, Faculty of Science, University of Amsterdam, Nieuwe Achtergracht 166, 1018 WVAmsterdam, The Netherlands.
Abstract:
Understanding and controlling the band broadening is essential to obtain accurate molar-mass distributions by size-exclusion chromatography (SEC). In this paper, band broadening in SEC is reviewed from a contemporary perspective. The observed band broadening is due to dispersion inside and outside the chromatographic column (undesirable band broadening) and to the polydispersity of the sample (desirable SEC selectivity). The various contributors to band broadening are discussed. Integrity plots are introduced as a tool to evaluate the performance of specific SEC columns at given experimental conditions. For narrow polymer standards on single SEC columns the observed peak width is dominated by the chromatographic dispersion. MALDI-ToF-MS is demonstrated as an alternative to determine the PDI of narrowly distributed samples. The plate heights encountered at very high reduced velocities are found to be lower than expected. This is advantageous for fast separations by SEC.
Related Concept Videos
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...