Related Experiment Video
Updated: Aug 20, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
GlucoWatch G2 Biographer alarm reliability during hypoglycemia in children
Eva Tsalikian1, Craig Kollman, Nelly Mauras
1DirecNet Coordinating Center, Jaeb Center for Health Research, Tampa, FL 33647, USA. direcnet@jaeb.org
Background:
The GlucoWatch G2 Biographer (GW2B) (Cygnus, Inc., Redwood City, CA) provides near-continuous monitoring of glucose values in near real time. This device is equipped with two types of alarms to detect hypoglycemia. The hypoglycemia alarm is triggered when the current glucose measurement falls below the level set by the user. The "down alert" alarm is triggered when extrapolation of the current glucose trend anticipates hypoglycemia to occur within the next 20 min.
Methods:
We used data from an inpatient accuracy study to assess the performance of these alarms. During a 24-h clinical research center stay, 89 children and adolescents with Type 1 diabetes mellitus (3.5-17.7 years old) wore 174 GW2B devices and had frequent serum glucose determinations during the day and night.
Results:
Sensitivity to detect hypoglycemia (reference glucose < or = 60 mg/dL) during an insulin-induced hypoglycemia test was 24% with the hypoglycemia alarm alone and 88% when combined with the down alert alarm. Overnight sensitivity from 11 p.m. to 6 a.m. was 23% with the hypoglycemia alarm alone and 77% when combined with the down alert alarm. For 16% of hypoglycemia alarms, the reference glucose was above 70 mg/dL for 30 min before and after the time of the alarm. For the two alarm types combined, the corresponding false-positive rate increased to 62%.
Conclusions:
The down alert alarm substantially improves the sensitivity of the GW2B to detect hypoglycemia at the price of a large increase in the false alarm rate. The utility of these alarms in the day-to-day management of children with diabetes remains to be determined.
Related Concept Videos
Hypoglycemia and Glucagon
Hypoglycemia
Hyperglycemia
Oral Hypoglycemic Agents: Glinides
Diabetes Mellitus: Type 2 and Gestational
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes: